510(k) Data Aggregation

    K Number
    K170290

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-24

    (114 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination in between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs:

    Chemotherapy Drug Permeation
    The following chemicals have been tested with these gloves.

    Chemotherapy Drug and ConcentrationBreakthrough Detection Time in Minutes
    Carmustine (BCNU), 3.3mg/ml27.8
    Cisplatin, 1.0mg/ml>240
    Cyclophosphamide (Cytoxan), 20mg/ml>240
    Cyclosporin A, 100 mg/ml>240
    Dacarbazine (DTIC) 10g mg/ml>240
    Doxorubicin Hydrochloride, 2.0 mg/ml>240
    Etoposide (Toposar), 20mg/ml>240
    Fluorouracil 50mg/ml>240
    Paclitaxel (Taxol), 6.0mg/ml>240
    Thiotepa, 10mg/ml77.9
    Vincristine Sulfate, 1.0mg/ml>240

    Please note that the following drug has extremely low permeation times: Carmustine (BCNU): 27.8 minutes.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA regarding a medical glove. It contains information about the device, its classification, and its intended use, particularly its resistance to chemotherapy drugs.

    However, the document does not contain the kind of information needed to answer your questions about acceptance criteria for a device performance study, nor does it describe a study that proves the device meets acceptance criteria in terms of AI model performance, human reader improvement with AI, or ground truth establishment.

    Specifically, the document lists specifications for the glove's resistance to various chemotherapy drugs (breakthrough detection time), which could be considered "acceptance criteria" for that specific characteristic. However, your detailed questions (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study, standalone performance, etc.) are characteristic of the evaluation of an Artificial Intelligence/Machine Learning (AI/ML) powered medical device. This document describes a traditional, non-AI medical device (a glove).

    Therefore, I cannot provide the requested information from this document. The information you are asking for is typically found in clinical study reports or detailed technical documentation for AI/ML medical devices, not in a 510(k) clearance letter for a patient examination glove.